3.156.
Following justification, the diagnostic radiological procedure or image guided interventional procedure is required to be performed in such a way as to optimize patient protection (para. 3.163 of GSR Part 3 [3]). The level of image quality sufficient for diagnosis is determined by the radiological medical practitioner and is based on the clinical question posed and the anatomical structures imaged (e.g. the diagnosis of the pattern of sinusitis on CT requires only a low dose procedure as high contrast structures, namely air and bone, be imaged). With image guided interventional procedures, the level of image quality should be sufficient to guide the intervention.
3.157.
The following points apply to all diagnostic radiological procedures or image guided interventional procedures:
There should be an effective system for correct identification of patients, with at least two, preferably three, forms of verification, for example name, date of birth, address and medical record number.
Patient details should be correctly recorded, such as age, sex, body mass, height, pregnancy status, current medications and allergies.
The clinical history of the patient should be reviewed.
3.158.
The first step in operational considerations of optimization is selection of the appropriate medical radiological equipment. For example, a chest X ray should be performed using dedicated equipment with a radiation generator producing high output enabling the use of a long source to image receptor distance (typically 1.8 m) and a short exposure time to ensure a reproducible image of diagnostic quality by minimizing patient respiratory motion and cardiac motion.
3.159
The volume (area) of the patient that is exposed should be strictly limited to that of clinical interest. This is achieved through collimation in radiography, mammography, fluoroscopy and image guided interventional procedures, and through the choice of scan parameters in CT. For diagnostic radiology, image cropping performed after the exposure does not achieve any reduction in the exposed volume.
3.160.
Cooperation of the patient should be ensured to achieve an image of diagnostic quality. This is particularly relevant when imaging children. Good communication helps to achieve this. Verbal interaction between the medical radiological technologist or the medical radiological practitioner and the patient should take place before, during and after the procedure.
3.161.
Optimization of protection and safety for a woman undergoing a radiological procedure during pregnancy should take into account the woman and the embryo or fetus. Routine diagnostic CT examinations of the pelvic region with and without contrast injection can lead to a dose of 50 mSv to the uterus, which is assumed to be the same as the dose that would be received by the fetus in early pregnancy. When CT scanning is indicated for a pregnant patient, low dose CT protocols should be used and the scanning area should be reduced to a minimum (see also paras 3.176–3.185).
3.162.
Shielding of radiosensitive organs, such as the gonads, the lens of the eye, the breast and the thyroid, should be used when appropriate. Care should be taken in the anatomical placement of such shields, the impact of shielding on image quality (artefacts), and the use of AEC devices and the consequences for patient dose.
3.163.
For each modality, there are a number of factors that can be adjusted to influence the relationship between image quality and patient dose. Written protocols that specify the operating parameters to be used for common diagnostic radiological procedures should be developed, adopted and applied in each radiology facility. Such protocol ‘technique charts’ should be posted adjacent to each X ray generator and should be specific for each piece of equipment. The protocols should take into account the anatomical region, as well as patient mass and size. The protocols should be developed using guidelines from national or international professional bodies, and hence should reflect current best practices (e.g. see Refs [137–147]). For modern digital equipment, many of the factors are automated through the menu driven selection of options on the console. Nevertheless, in setting up these options, significant scope exists for the optimization of protection and safety through the appropriate selection of values for the various technical parameters, thereby effectively creating an electronic technique chart.
3.164.
Size specific written protocols should be developed for children, from neonates to teenagers, and should include additional operational considerations, such as the use of additional filtration or the removal of grids when appropriate [143, 145, 146].
3.165.
Paragraph 3.166(b) of GSR Part 3 [3] establishes a special requirement for the optimization of protection and safety for individuals subject to medical exposure as part of an approved health screening programme. All aspects of protection should be considered before the approval of the programme and during its implementation, such as the selection of X ray equipment suitable for the particular screening and parameters settings. A dedicated, comprehensive programme of quality assurance should be implemented to meet screening objectives, as described in more detail in paras 3.232–3.246. It should set requirements for the education and training of the medical professionals involved in the health screening programme, for adequate quality management for the whole screening chain and for documentation and evaluation of the results.